Halving the Volume of AnaConDa: Evaluation of a New Small-Volume Anesthetic Reflector in a Test Lung Model

BACKGROUND: Volatile anesthetics are increasingly used for sedation in intensive care units. The most common administration system is AnaConDa-100 mL (ACD-100; Sedana Medical, Uppsala, Sweden), which reflects volatile anesthetics in open ventilation circuits. AnaConDa-50 mL (ACD-50) is a new device with half the volumetric dead space. Carbon dioxide (CO2) can be retained with both devices. We therefore compared the CO2 elimination and isoflurane reflection efficiency of both devices. METHODS: A test lung constantly insufflated with CO2 was ventilated with a tidal volume of 500 mL at 10 breaths/min. End-tidal CO2 (EtCO2) partial pressure was measured using 3 different devices: a heat-and-moisture exchanger (HME, 35 mL), ACD-100, and ACD-50 under 4 different experimental conditions: ambient temperature pressure (ATP), body temperature pressure saturated (BTPS) conditions, BTPS with 0.4 Vol% isoflurane (ISO-0.4), and BTPS with 1.2 Vol% isoflurane. Fifty breaths were recorded at 3 time points (n = 150) for each device and each condition. To determine device dead space, we adjusted the tidal volume to maintain normocapnia (n = 3), for each device. Thereafter, we determined reflection efficiency by measuring isoflurane concentrations at infusion rates varying from 0.5 to 20 mL/h (n = 3), for each device. RESULTS: EtCO2 was consistently greater with ACD-100 than with ACD-50 and HME (ISO-0.4, mean ± standard deviations: ACD-100, 52.4 ± 0.8; ACD-50, 44.4 ± 0.8; HME, 40.1 ± 0.4 mm Hg; differences of means of EtCO2 [respective 95% confidence intervals]: ACD-100 − ACD-50, 8.0 [7.9–8.1] mm Hg, P

from Anaesthesiology via xlomafota13 on Inoreader https://ift.tt/2IFlOft
via IFTTT

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου

Σημείωση: Μόνο ένα μέλος αυτού του ιστολογίου μπορεί να αναρτήσει σχόλιο.