Abstract
Background
Endoscopic irreversible electroporation (IRE) can be performed using a flexible, thin, needle-shaped electrode for an endoscopic ultrasound (EUS)-guided procedure. This study aimed to evaluate the feasibility and efficacy of performing EUS-guided IRE with endoscopic needle-electrode in porcine pancreas.
Methods
Experimental endoscopic IRE on the pancreas were performed by EUS-guided approach in three pigs and compared with surgical approach in three pigs. The animals were killed after 24 h and their pancreases collected.
Results
IRE ablation using endoscopic needle-electrode was successful technically in EUS-guided approaches for the pancreas. Immediately following IRE, the ablated pancreatic tissue showed no gross change except focal hemorrhage. H&E staining presented a well-demarcated ablation site measuring 1.0–1.5 cm in diameter in the pancreas. TUNEL immunohistochemistry showed diffuse cell death along the puncture site 24 h after IRE. No complication was observed in pigs after endoscopic IRE ablation.
Conclusion
EUS-guided IRE ablation was feasible and effective for pancreas using the newly developed device.
from Endoscopy via xlomafota13 on Inoreader http://bit.ly/2FP63F3