Παρασκευή 3 Νοεμβρίου 2017

Control of inferior vena cava injury during laparoscopic surgery using a double balloon-equipped central venous catheter: proof of concept in a live porcine model

Abstract

Background

Iatrogenic inferior vena cava (IVC) injury is a rare but potentially life-threatening complication during laparoscopic surgery. This experimental study aimed to assess the hemostatic ability of a new device, double balloon-equipped central venous (DB-CV) catheter, for IVC injury.

Methods

The DB-CV catheter comprises a triple-lumen sphincterotome combined with two dilating balloons having a diameter of 25 mm. The experimental procedures were performed in five pigs. The DB-CV catheter was inserted via the right femoral vein. For the IVC occlusion test, correct placement of the balloons was confirmed by indocyanine green fluorescence imaging, and hemodynamic data were recorded. For the IVC injury test, a 3- to 4-mm circumferential incision was created in IVC, and hemostasis was initiated using balloon inflation 5 s after the injury.

Results

Hemodynamic changes were minimal, with a 20 mmHg reduction in the mean arterial pressure because of IVC occlusion. All bleeding from IVC injuries was successfully temporarily stopped by direct balloon compression, with a mean time to hemostasis of 69 s and mean blood loss of 32 ml. Subsequently, the positioning of IVC injuries between two balloons made it possible to suture the injured IVC.

Conclusions

Balloon occlusion using the DB-CV catheter provides a rapid temporal hemostatic effect and can overcome the serious condition of massive hemorrhage from IVC injuries.



from Endoscopy via xlomafota13 on Inoreader http://ift.tt/2iYIcIS
via IFTTT

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου

Σημείωση: Μόνο ένα μέλος αυτού του ιστολογίου μπορεί να αναρτήσει σχόλιο.