Abstract
Background
Advanced virtual monochromatic reconstruction from dual-energy brain CT has not been evaluated in children.
Objective
To determine the most effective advanced virtual monochromatic imaging energy level for maximizing pediatric brain parenchymal image quality in dual-energy unenhanced brain CT and to compare this technique with conventional monochromatic reconstruction and polychromatic scanning.
Materials and methods
Using both conventional (Mono) and advanced monochromatic reconstruction (Mono+) techniques, we retrospectively reconstructed 13 virtual monochromatic imaging energy levels from 40 keV to 100 keV in 5-keV increments from dual-source, dual-energy unenhanced brain CT scans obtained in 23 children. We analyzed gray and white matter noise ratios, signal-to-noise ratios and contrast-to-noise ratio, and posterior fossa artifact. We chose the optimal mono-energetic levels and compared them with conventional CT.
Results
For Mono+maximum optima were observed at 60 keV, and minimum posterior fossa artifact at 70 keV. For Mono, optima were at 65–70 keV, with minimum posterior fossa artifact at 75 keV. Mono+ was superior to Mono and to polychromatic CT for image-quality measures. Subjective analysis rated Mono+superior to other image sets.
Conclusion
Optimal virtual monochromatic imaging using Mono+ algorithm demonstrated better image quality for gray–white matter differentiation and reduction of the artifact in the posterior fossa.