Assessment of a modified technique for air delivery during fluoroscopic-monitored pneumatic intussusception reduction

Abstract

Background

Intussusception is one of the most common causes of acute abdominal emergency in children. Image-monitored pressure reduction is the first line of treatment.

Objective

We report on a modified technique of air delivery during fluoroscopic-monitored pneumatic intussusception reductions, and compare it with an established technique.

Materials and methods

We modified the Shiels intussusception reduction device so that the air used for intussusception reduction is delivered not by the device’s insufflator bulb, but rather by the hospital medical air supply system, eliminating the need for continuous pumping of the insufflator bulb during the procedure. Subsequently, we retrospectively compared sequential fluoroscopy-monitored pneumatic intussusception reduction procedures performed in patients younger than 18 years using either the standard or modified devices, evaluating technical procedure parameters (i.e. median procedure time length, fluoroscopy time length and radiation dose) and patient outcomes (i.e. number of complete intussusception reductions, number of incomplete intussusception reductions, number of intussusception recurrences within 3 days of the procedure, number of procedures followed by surgery).

Results

We found no statistically significant differences between procedures performed with the standard and modified techniques.

Conclusion

The device modification allows for increased operator comfort. Evaluated procedure parameters and patient outcomes appear similar to those of the standard technique.