Abstract
Purpose
The fourth-generation FloTrac/Vigileo™ improved its algorithm to follow changes in systemic vascular resistance index (SVRI). This revision may improve the accuracy and trending ability of CI even in patients who undergo abdominal aortic aneurysm (AAA) surgery which cause drastic change of SVRI by aortic clamping. The purpose of this study is to elucidate the accuracy and trending ability of the fourth-generation FloTrac/Vigileo™ in patients with AAA surgery by comparing the FloTrac/Vigileo™-derived CI (CIFT) with that measured by three-dimensional echocardiography (CI3D).
Methods
Twenty-six patients undergoing elective AAA surgery were included in this study. CIFT and CI3D were determined simultaneously in eight points including before and after aortic clamp. We used CI3D as the reference method.
Results
In the Bland–Altman analysis, CIFT had a wide limit of agreement with CI3D showing a percentage error of 46.7%. Subgroup analysis showed that the percentage error between CO3D and COFT was 56.3% in patients with cardiac index < 2.5 L/min/m2 and 28.4% in patients with cardiac index ≥ 2.5 L/min/m2. SVRI was significantly higher in patients with cardiac index < 2.5 L/min/m2 (1703 ± 330 vs. 2757 ± 798; p < 0.001). The tracking ability of fourth generation of FloTrac/Vigileo™ after aortic clamp was not clinically acceptable (26.9%).
Conclusions
The degree of accuracy of the fourth-generation FloTrac/Vigileo™ in patients with AAA surgery was not acceptable. The tracking ability of the fourth-generation FloTrac/Vigileo™ after aortic clamp was below the acceptable limit.
from Anaesthesiology via xlomafota13 on Inoreader https://ift.tt/2Gu2twG
via IFTTT