Τετάρτη 25 Ιουλίου 2018

Do Epidural Catheter Size and Flow Rate Affect Bolus Injection Pressure in Different Programmed Intermittent Epidural Bolus Regimens? An In Vitro Study

BACKGROUND: The optimal programmed intermittent epidural bolus regimen for labor analgesia remains unknown. Some studies indicate that better drug spread in the epidural space results from greater injection pressure; however, there is a lack of data regarding the maximum pressure generated by epidural bolus injection using different catheters and flow rates. METHODS: We evaluated the flow and pressure characteristics of 11 commonly used epidural catheters combined with 3 different infusion pumps that deliver epidural infusions according to the programmed intermittent epidural bolus regimen. Pressure changes were measured over time at flow rates of 100, 250, and 400 mL·hour−1 and with a bolus volume of 10 mL. To account for repeated measures, linear mixed models were used. Features were selected with a backward stepwise procedure continued until only statistically significant variables were left in the model. RESULTS: We performed 660 measurements. The mean maximal pressure generated during bolus injection ranged from 86 to 863 mm Hg for different flow rates and catheter designs. The interaction between flow rate and catheter gauge resulted in 1.31, 1.65, and 2.00 mm Hg of pressure increase for 18G, 19G, and 20G catheters, respectively, per 1 mL·hour−1 of increased flow rate (P

from Anaesthesiology via xlomafota13 on Inoreader https://ift.tt/2v3JmFn
via IFTTT

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου

Σημείωση: Μόνο ένα μέλος αυτού του ιστολογίου μπορεί να αναρτήσει σχόλιο.