Τρίτη 26 Απριλίου 2016

Comparison of Changes in Tensile Strength in 3 Different Flexible Epidural Catheters Under Various Conditions.

BACKGROUND: Rarely, epidural catheters may fracture upon removal or insertion. Understanding some of the mechanical properties of epidural catheters, such as their tensile strength and how external factors (including temperature) can influence their strength, will aid physicians in making decisions if faced with an entrapped catheter. In the present study, we evaluated the impact in tensile strength when catheters are exposed to 37 +/- 1[degrees]C, after the removal of the inner metal coil and after the injection of sterile saline through the catheter. METHODS: We analyzed the tensile strength of a total of 120 catheters (19-gauge) from 3 different brands and materials. The reinforced epidural catheters were affixed to opposing, specially designed tensile test fixtures and then installed in an Applied Test System tensile test apparatus. We evaluated the strength of 10 catheters from different brands and materials for each of the following variables: at room temperature (control group), after the removal of the inner wire present in all the flexible catheters tested, injection of normal saline, and at 37 +/- 1[degrees]C. RESULTS: When compared with their control groups, the Arrow catheter (2.85 kg) was shown to be superior to B-Braun (2.17 kg; P

from Anaesthesiology via xlomafota13 on Inoreader http://ift.tt/1SLfUfZ
via IFTTT

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου

Σημείωση: Μόνο ένα μέλος αυτού του ιστολογίου μπορεί να αναρτήσει σχόλιο.